Gaseous Applications-

Products and Components

Shorten your cycle times by using Clordisys.

Lumen Sterilizer Load (Total Sterilization Cycle Time - 3 hour 40 minutes)

Cycle Parameters:

Note: No External Aeration Required

Chamber Volume (cu ft):

21

Vacuum Sp (K Pa):

15

Leak Test Sp (K Pa):

30

Leak Test Time (min):

1

Leak Test Allowable Drop (Pa):

10

 

 

RH Condition Sp (%RH):

70

 RH Condition Time (min):

30

CD Concentration Sp (mg/L):

30

CD Expose Time (min):

60

Vacuum / Break Cycles:

3

Aeration Control:

Manual

 

Lumen Sterilizer Load Chart

setstats 1