Steridox-VP
The Steridox-VP CD Sterilizers are designed for use in any pharmaceutical, manufacturing, laboratory, research or surgical setting. They provide a rapid and highly effective method to sterilize medical devices, sterile products, instruments, and components at ambient temperatures. The sterilizer utilizes a rugged industrial control system. It features a sophisticated sterilant concentration monitoring system to assure a tightly controlled sterilization process facilitating parametric release. All instrumentation, including the photometer for concentration monitoring, is easily calibrated to traceable standards. The process is easy to validate due to the repeatable cycle, tight process control, and highly accurate sterilant monitoring system. A run record is produced that contains: date, cycle time, cycle steps, as well as temperature, pressure, and chlorine dioxide concentration. The sterilizer is available in a variety of sizes to meet your processing needs. Automated sliding doors are available. The chamber and doors are constructed of 316L stainless steel to provide an exceptionally long life. The unit is enclosed in stainless steel, easily removable panels. The process is non-explosive allowing for inexpensive installations.
Why use the Steridox-VP?
|
·
Sterilization
at ambient temperatures ·
short
cycle times ·
precise
concentration monitoring ·
Uses
a true gas ·
excellent
distribution into hard to reach areas ·
quick
aeration (can literally be minutes) ·
gas
concentration feedback to control system ·
no
down time between cycles ·
does
not require tight control of dew point temperatures ·
does
not condense out or breakdown during the process ·
extremely
low residuals · rugged and reliable industrial components · quality construction |
·
simple
to validate ·
complete
cycle automation ·
easy
to use touch screen operator interface ·
detailed
cycle reporting ·
recipe
management ·
historical
and real time trending ·
no
liquids in process ·
quick
consumable replacement ·
rapid
aeration when compared to ethylene oxide ·
no
concern for condensation as with hydrogen peroxide ·
non-carcinogenic ·
non-flammable · no need for Damage Limiting Construction (DLC) |
Options:
| Chamber Size: | Standard Sizes: 16” x 16” x 26”D, 20” x 20” x 38”D;
26” x 26” x 39”D
Custom sizes up to 1000 cubic feet |
| Door Style: | Single or Double Power Operated Door |
| Humidification Steam: | House Supplied or Integral with Sterilizer |
| Fascia: | Recessed (flange) style or Cabinet style |
| Rack/Loading options: | Multi-position Rack; Extendable Shelves;
Single Shelf; Loading Car; Transfer Carriage
|
| Multi-process: | Optional units that can sterilize by either steam or CD are available |